Model 6040
TTL/ECL Pulse & Delay Generator
Specifications
Operating Capabilities
- Single Pulse
- Conventional pulse generator operation with selections
for rate, amplitude, width and delay.
- Double Pulse
- Pair of identical single pulses with leading edges
separated by a selected delay.
- Digital Delay
- Selectable duration between an external trigger or
mainframe Trig(sync) output and the pulse outputs of the
mainframe or a module.
- Single Impulse
- Subnanosecond fixed width optical "spikes" with
selections for rate and delay.
- Double Impulse
- Pair of identical optical impulses separated by a
selected delay.
- External Drive
- Pulse output corresponds to rate and timing of an
external input pulse train, sine waves, etc. Also can
allow operation at rates well above 100 MHz.
- External Modulation
- Analog or digital signals are converted into their
optical equivalents. The input modulates light from a
selectable quiescent level.
- CW
- An adjustable steady-state power level of light.
Timing
- Internal Trigger
- 0.01 Hz to 100 MHz.
Accuracy: 0.01% of setting.
- External Trigger
- Trig In: 0 to 100 MHz.
Threshold Range: +/- 2.5 V.
Threshold Resolution: 10 mV.
Slope: Selectable positive or negative edge.
Sensitivity: 100 mV peak-to-peak for widths >1 ns; 200
mV for 500 ps width.
Input Impedance: 50 ohms.
- Pulse Width
- 3ns to 640s.
Resolution: lns or 5 digits, whichever is greater.
Jitter: 25ps rms or 0.005% x width, whichever is greater.
Accuracy: Up to 9ns- +/ lns; lOns or more- +/ 500ps or
0.2% x width, whichever is greater.
- Delay
- Variable: 0 ns to 640 s. Fixed(typical): Trig In to Trig
Out 20 ns; Trig Out or ECL out 30 ns.
Resolution: 1 ns or 5 digits, whichever is greater.
Jitter: 25ps rms or 0.005% x delay, whichever is greater.
Accuracy: 500 ps or 0.2% x delay, whichever is greater.
Outputs
- Trig Out
- A time-zero sync pulse at the start of each internal
operating cycle.
Amplitude: 2 V into 50 Ohms.
Transition Times: 3 ns rise time, 3 ns fall time.
Width: 3 ns.
Trigger Delay: 20 ns typical(fixed) Trig In to Trig Out.
Jitter: 25 ps rms.
- Pulse Out
- Mainframe front panel TTL-compatible output controllable
through the full ranges of repetition rate, pulse width,
delay or double pulse timing.
Amplitude: 5 V into 50 Ohms.
Transition Times: 1 ns rise time, 1.5 ns fall time.
Insertion Delay: 50 ns typical(fixed) Trig In to Pulse
Out. Jitter: 30 ps rms.
- ECL Out
- Mainframe rear panel ECL-compatible output coincident
with and controlled as for Pulse Out.
Amplitude: -0.8 V to -1.8 V into 50 Ohms.
Transition Times: 700 ps rise time, 700 ps fall time.
Insertion Delay: 50 ns typical(fixed) Trig In to ECL Out.
Jitter: 30 ps rms.
Programming
- IEEE-488
- Remote interface with all functions and parameters
programmable and bus triggerable. Interface Functions:
SH1, AH1, TT6, TE0, SR1, RL1, PP0, DC1, DT1, C0.
(IEEE-488.2-1987)
- RS-232C
- Remote interface with all functions and parameters
programmable in full- or half-duplex to 1.2 kbaud and bus
triggerable.
- Local
- Pushbutton manual entry with panel LCD display.
www.r-systems.com