SPECIFICATIONS Model 4800
Measurement: Range and Accuracy
Gas Range Accuracy
CO2 0.0 to 30 Torr ± 1.5 Torr

30.1 to 76 Torr    5% Relative
N2O 0.0 to 100% ± 1.5% abs. plus 5% rel.
Halothane 0.00 to 7.5% ± 0.1% abs. plus 4% rel.
Isoflurane 0.00 to 7.5% ± 0.1% abs. plus 4% rel.
Enflurane 0.00 to 7.5% ± 0.1% abs. plus 4% rel.
Sevoflurane 0.00 to 9% ± 0.1% abs. plus 4% rel.
Desflurane 0.00 to 20% ± 0.1% abs. plus 4% rel.
Halocarbon 23 0.00 to 10% ± 0.4% abs.
Oxygen 0.00 to 100% ± 3.0% abs.
Ethanol 0.0 to 0.1% ± 0.013% abs.
Agent Identification
Detection Time 10 sec.(max)
Primary Agent Identification Sensitivity
Primary Agent Identified Primary Agent Threshold
Isoflurane 0.20% Isoflurane abs.
Halothane 0.20% Halothane abs.
Enflurane 0.20% Enflurane abs.
Sevoflurane 0.24% Sevoflurane abs.
Desflurane 0.30% Desflurane abs.
Secondary Agent Identification Sensitivity
Secondary Agent %Absolute of Secondary Agent %Relative of Primary Agent
Halothane 0.20% (abs.) Halothane ±16.0% (rel.) Des., Sev., Iso., Enf.
Isoflurane 0.50% (abs.) Isoflurane ±5.0% (rel.) Hal., Sev.
±6.0% (rel) Enf.
±18.0% (rel) Des.
Enflurane 0.20% (abs.) Enflurane ±5.0% (rel.) Iso., Hal.
±6.0% (rel) Des.
±10.0% (rel.) Sev.
Sevoflurane 0.24% (abs.) Sevoflurane ±5.0% (rel.) Iso., Hal.,Enf.
±6.0% (rel.) Des.
Desflurane 0.30% (abs.) Desflurane ±8.0% (rel.) Iso., Hal., Enf., Sev.
Example: Take the case where the primary agent is 2.00% Halothane. The maximum 
Enflurane concentration required to identify it as a secondary agent is 0.30% Enflurane
(0.20%+(0.05x2.00)).
Acetone: on command
Alcohol: automatic, real time
Warm-Up Time
0 to 60 seconds no data reported
60 seconds (typically) full accuracy for all gases
Response Time Specifications [T10 to T90]
CO2 300 millisecs.
N2O 450 millisecs.
Ethanol 700 millisecs.
All agents 700 millisecs.
O2 (fast) 300 millisecs.
Breath Rate Data
Breath Rate: 0 to 100bpm ± 2bpm
General
Size (WxDxH)  151x376x93 mm / 5.95x14.80x3.65 in.
Weight  3.85 kg / 8.5 lbs.
Sampling Gas Flow Rate 120 & 200 ml / min.
Power  AC   DC
Input Voltage  100 to 240 Vac + 10%  19 to 30 Vdc
(90 to 264 Vac)
Frequency 48 to 62 Hz
Power Consumption  22 watts typical /
36 watts peak
Communications  RS232C asynchronous
serial, 19,200bps
Calibration
Span  Not required. Span stable for life.
Zero <10 sec.
Flash Memory Reprogramable Flash Memory upgrades via modem
Operating Environment
Temperature  +10 to +40°C / +50 to 104°F
Humidity  10 to 95% RH (non-condensing)
Pressure/Altitude 525 Torr to 790 Torr, (1060hPa to 700hPa)
/ -305 m to + 3,048 m
Regulatory Approvals
UL 2601-1  IEC 60601-1
CSA-C22.2 IEC 1000-4-4
IEC 801-2 IEC 1000-4-5
IEC 801-3 CISPR 11.1 & 22
EN 60601-1 FCC Class B

Anesthesia Gas Signatures
-Halothane -Enflurane -Isoflurane -Desflurane -Sevoflurane